Module 1 Section 2

 

Home

Catalog

Go to Section 3

       
       
 

Module 1:

Basic A-scan Biometry
 

Section 2:

Scanning Procedure
     
 

Introduction

Preparation

Hand-held method

Tonometer mount method

Immersion method

Probe positioning

Scan evaluation

Aphakic scan

Measuring gates

Extra lens spikes

Corneal compression

Reproducibility

 
   
 

Introduction

There are three basic methods which are used to obtain an axial length measurement using an A-scan ultrasound instrument.  Each method has its pros and cons.  No matter which method is used,  there are techniques to be used and scan characteristics to watch for that can improve the accuracy of the scan.

Preparation

Instrument set-up:  The A-scan instrument is prepared for measurement.  Some instruments require a calibration procedure before a measurement is performed. 

Mode selection:  Most instruments require you to select a mode of measurement.  This insures that the instrument is using the correct velocity-of-sound for the particular eye you are measuring.  The choices are usually:

 

  • Phakic (cataract, or cataractous) - This is usually the default choice because it is most often used.
  • Dense cataract - A dense cataract may absorb so much ultrasound energy that there is little energy left to produce a high retina spike.  This mode usually increases the gain level in the automatic mode.
  • Aphakic (no lens) - This is usually used to measure an eye that will have an IOL inserted after a previous surgery in which the lens was removed but no IOL was inserted.
  • Pseudophakic (IOL in place) - This is used to measure an eye that already has an IOL in place.

Patient instruction:  An informed patient will be more cooperative.  Tell the patient that you will be performing some measurements to determine the power of the implant that will replace his cataractous lens.  Tell the patient that you may be measuring both eyes for comparison.  Some patients get upset if you start by measuring the eye that is not going to have surgery.

Keratometry:  Perform keratometry before you perform A-scan biometry,  and before any drops are instilled.  This will eliminate any disruption of the corneal surface.  Lubricating drops can be used during keratometry if the mires are not clear.

Disinfection:  A solid probe tip should be disinfected with an alcohol wipe before use.  Wipe the tip or let it air dry before it is used.   If there is some question about contamination, a 10 minute soak in 3% hydrogen peroxide can be used.  A solution of 1 part household bleach to 10 parts distilled water is also an effective disinfectant after a 10 minute soak.

Membrane type tips should be rinsed with saline before use.  If there is some question about contamination, a new tip should be used.  Membrane type tips are not used on new instruments, so they are becoming obsolete.

Hand-held method

The probe is hand held by the examiner and is applied to the cornea without any other means of support.  The patient can be instructed to fixate on the light inside the probe tip.  A fixation target for the fellow eye, such as an X on the wall, can also be used.  If necessary, the patient can be simple instructed to stare straight ahead.

Advantages:  No additional set-up is required.   There is good control over patient positioning.  There is less dependence on steady patient fixation.  The patient does not have to maintain positioning in a headrest, and the examiner can easily hold the eyelids open if necessary.

Disadvantages:  The examiner has less precise control over the positioning of the probe relative to the eye.  There is a danger of putting too much pressure on the cornea, causing corneal compression and a shortening of the axial eye length.  This is a significant potential error source.  A .4mm compression error can result in a 1 diopter error in the calculated IOL power.

Tonometer mount method

The A-scan probe is mounted in place of the prism tip on the applanation tonometer.  The patient is positioned at the slit-lamp and the probe is applied to the cornea in the same fashion that applanation tonometry is performed.  The patient is instructed to fixate on the light in the probe, or the slit-lamp fixation light can be used with the fellow eye.

Remove this screw.Set-up:  Some probes will not fit into tonometer prism holder if the screw is in place.  You can remove the small screw from the prism holder if it is still in place.  This screw is meant to prevent the tonometer prism from being inserted backwards, but it may also prevent you from mounting the A-scan probe.

 

 

 

Tonometer mounted probe.

Insert the A-scan probe into the tonometer prism holder with the tip pointing toward the patient.  Loop the cord of the probe in such a manner as there is no tension on the cord. 

Decrease the reading on the pressure drum until the probe tips backwards.  Now increase the drum reading until the probe just tips forward.   This will insure that there is minimal pressure on the cornea.

 

 

Advantages:  A cooperative patient can be held in a steady position.  The slit lamp allows more precise control of the probe.  The forward pressure of the probe tip is somewhat limited by the tonometer mount,  thus minimizing the possibility of corneal compression by the probe tip.

Disadvantages:  Additional set-up is required.   Good patient fixation is critical because the probe cannot be adjusted on the vertical axis (the tonometer mount does not allow the probe to be tipped up or down).   The patient must maintain position in the head rest.  It is more difficult to assist patients who cannot keep the eye open.

Immersion method

Immersion A-scanThe immersion method is employed to eliminate corneal compression as an error factor in axial length measurements.  A "shell" is placed on the eye to provide a "water bath" over the cornea.  The water bath provides a medium through which the probe can measure the eye without touching the cornea.  For a detailed discussion of the technique, see Module 3, Section 2.

 

 

Probe positioning

Alignment with the optical axis.The probe lightly touches the cornea and is positioned such that the barrel of the probe is aligned with the optical axis of the eye.  The operator aims the probe toward the macula of the eye.  Alignment is guided by the appearance of the scan or by some other signal from the A-scan instrument.  Alignment with the optical axis will be indicated by high lens spikes and a high retina spike on the scan graph.

 

 

Freezing a good scan.Spike height will vary as small adjustments are made in the alignment of the probe relative to the cornea.  In the manual mode the operator will have to freeze the scan with a foot-pedal when an acceptable scan is obtained. The gain should be adjusted high enough such that the spikes can be maintained above the threshold level needed for an automated acquisition of the scan if this feature is used.   The gain should be low enough to allow the operator to visually maximize the spike height during probe alignment.

 

Scan evaluation

For an A-scan to be an acceptable measurement of the axial length of a phakic eye, it must meet certain criteria:

 

  • Characteristics of a good A-scan.The anterior lens spike (B), the posterior lens spike (C), and the retina spike (E) must be present. The probe tip / cornea spike is represented by (A).
  • These spikes should be tall and steeply rising.  The retina spike should not have smaller spikes immediately in front of it (D).
  • The retina spike should be followed by a tall scleral spike (F) and spikes from the orbital fat layer of the orbit (G).  A scan without orbital fat spikes may indicate that the beam is striking the optic nerve instead of the macula.

 

What's wrong with this scan?

What's wrong with this scan?

1) A lens spike is missing 2) There is a small spike in front of the retina spike 3)  The orbital fat spikes are poorly defined.

 

 

Aphakic scanThe scan of an aphakic eye will either have no lens spikes, or it will have one lens spike (A) that represents an intact posterior lens capsule (C).   Be sure to use the aphakic mode of the A-scan instrument.  The velocity of sound will be different because the beam is not passing through the lens. A velocity of sound of 1532 m/s is typically used for aphakic measurements.  The instrument will automatically adjust for this if it has an aphakic measuring mode.

See Module 2:  Advanced A-scan Biometry for a discussion of the pseudophakic scan.

Measuring gates

A-scan instruments calculate measurements by means of electronic measuring "gates" or "lights".  To be sure that the instrument is obtaining a proper measurement, check the position of the gates on the display.

Measuring gates.This instrument has two gates: one for the cornea spike and  one that the retina spike must come within.  The gates are depicted in yellow at the bottom of the scan.

 

 

 

Measuring lights.

 

The instrument to the right has measuring lights that attach to the cornea, lens, and retina spikes.

 

 

Extra lens spikes.Extra lens spikes:  Sometimes cataract formation can result in irregular interfaces within the lens that create extra lens spikes.   These extra spikes can "fool" some instruments into attaching lights to one or more of these extra spikes.  This will result in a very short measurement because the total length is figured from the first light to the last light.

 

Corneal compression

Corneal compression occurs when too much pressure is exerted onto the cornea by the probe.  This can result in an erroneous axial eye length measurement.  Corneal compression can be monitored by observing the anterior chamber depth (AC, or ACD) read-out that most A-scan instruments provide.

Typically, an instrument will give a read-out for anterior chamber depth, lens thickness, and total eye length.  Most eyes will have an ACD reading between 2.5 and 4 mm.  Pay attention to the ACD reading on successive measurements.  If the ACD reads 3.1, 3.2, 3.1 on three measurements, and then reads 2.7 on the fourth, assume that the 2.7 results from a compressed cornea and throw out the measurement.

The corneal compression error factor can be avoided by using the immersion technique.  For more information see Module 2: Advanced A-scan Biometry.

Reproducibility

Reproducibility is an indicator of accuracy.   If you are to have confidence in the accuracy of your measurement, you must be able to reproduce the measurement.  This means being able to take four or five consecutive measurements with consistent appearance, consistent ACD read-outs,  and consistent total length results.  A commonly accepted level of reproducibility in terms of total length is to have several measurements within .20 mm of each other.

Top of page