Module 29 

 

Home

Catalog

Contact us

 

 

Get Post-Test

 

     
 

Module 29:

EOM Related Skills

 

Section 5:

Exophthalmometry 
   
   
 

 

The normal eye protrudes a distance of 16-20 mm as measured from the lateral orbital rim to the corneal apex. This distance is normally symmetrical with respect to the fellow eye.

 

 

Abnormal protrusion of one or both eyes is termed exophthalmos, or proptosis. A difference of more than 2 mm between eyes indicates unilateral exophthalmos. If an eye is thought to be shrinking back into the socket, it is termed enophthalmos (sometimes termed endophthalmos).

Exophthalmos can be caused by thyroid eye disease, an orbital tumor, air or blood in the orbit, and other less common causes.

Congenitally abnormal orbits, lids, and/or globes (axial myopia) can cause the appearance of exophthalmos without an underlying disease process. This is sometimes referred to as pseudoexophthalmos.

Exophthalmos is measured clinically using an exophthalmometer, typically a Hertel type exophthalmometer. Some simply refer to the instrument as "the Hertel" to avoid sounding like Elmer Fudd when trying to pronounce "exophthalmometer".

In order to get an accurate measurement of proptosis, we need a fixed point from which to measure. That fixed point is the lateral bony rim of the orbit, at the outer canthus. We measure from this point to the apex of the cornea.

The amount of proptosis may change over time, from one office visit to the next. These changes may be an indication of a change in the disease process. It is important to make sequential measurements in a manner that avoids errors due to variations in technique.

The instrument consists of two yokes mounted on a horizontal bar. Each yoke has a notch for alignment with the bony orbital rim, and a mirror, or mirrors, for observation of the corneal apex. One of the yokes is adjustable along the bar to accommodate the distance between the outer orbital rims of the two eyes. 

 

 

 

Procedure

 

Sit face to face and at eye level with the patient.

Hold the instrument so that the bar is toward you, and the mirrors are on top.

Hold the left side yoke in your left hand and the right side yoke in your right hand.

If there is a previous measurement to refer to, set the baseline indicator on the same mm mark. If you are using a B&L instrument, tighten the locking screw.

 

The exophthalmometer can be a real torture device if this step is not done properly. While holding the instrument in front of the patient’s eyes, locate the lateral orbital rims with your fingers before you place the notches on the bones. (Locate this orbital rim on your own eyes with your fingers. Notice that there isn’t much room between this bony rim and your eyeball. If you don’t make a good placement of the instrument, you may be putting uncomfortable pressure on the globe.) Keep your fingers on the orbital rims and guide the notches into contact with the rims.

 

After properly placing the instrument on the bones, have the patient look straight ahead. When measuring the patient’s left eye, close your left eye and correctly align your right eye with the mirror of the right yoke.

 

An image of the patient's cornea is reflected in the mirror. 

There is a mm scale on the front of the mirror housing. It is your job to read the mm mark that is closest to the image of the apex of the patient’s cornea.

Your line of sight must be perpendicular with the instrument in order for you to get an accurate measurement. There are two red marks in the mirror housing of the yoke that aid you in perpendicular alignment.

As you look into the mirror, observe the positions of the two red marks.  If you see both red marks, as pictured below, then you are not in perpendicular alignment.  You must then move your line of sight so that the two red marks superimpose.

In the photo below, the red marks are close to being superimposed.

 

When you are properly aligned, observe the image of the apex of the cornea and read the mm mark that is closest to the apex.

Record the measurement for each eye along with the baseline measurement. For example:

                               OD 22mm OS 23mm

                                     base: 102 mm

 

Sources of error

 

There are three potential sources of error when using the exophthalmometer.

The patient’s gaze: Make sure the patient is looking level and straight ahead for proper alignment of the corneal apex.

Instrument alignment: Make sure the notches are touching the orbital rims at the outer canthus, and that the instrument is level. If you have a baseline from a previous measurement, use the same baseline for this measurement.

The operator’s perspective: Make sure you are viewing from the correct angle (perpendicular). With the B&L instrument the zero mark should be just visible at the inside edge. With the Marco instrument the red lines should be superimposed, as discussed above.

For the B&L instrument, the correct alignment is pictured in the top graphic below.  Incorrect alignment is pictured in the bottom two graphics:

With the Marco instrument, the corneal image is superimposed onto the mm scale.  

With the B&L instrument, the image of the cornea is below the image of the scale.

 

   
  Back to top                   Get Post-Test